Parts & Service ECA 450
74 Instruction 0024-9400
5.4 Troubleshooting Guide
The following table lists the causes and remedies for most of the problems
that may arise with the analyzer.
If the information in Table 5-1 does not solve the problem, or for help with
any problem that is not listed, then please contact one of the Bacharach
Service Centers listed in Section 6.3.
TABLE 5-1. TROUBLESHOOTING GUIDE
Symptom
Probable Cause & Remedy
Analyzer completely nonfunctional;
won’t turn on when the I/O key is
pressed.
a. Battery pack dead. Recharge
battery pack per Section 4.1.
b. Loose battery connector. Remove
analyzer’s top panel and ensure that
the battery connector is securely
attached to the printed circuit
board.
c. Microprocessor needs reset.
Remove analyzer’s top panel and
press the Reset button shown in
Figure 4-5.
Display screen is blank when
analyzer is turned on, but pump runs
during warm-up cycle.
a. Microprocessor needs reset.
Remove analyzer’s top panel and
press the Reset button shown in
Figure 4-6.
b. Loose display connector(s).
Remove analyzer’s top panel and
ensure that all display-panel
connectors are properly attached to
the printed circuit boards.
Display goes blank during operation.
The analyzer’s display blanking
function has been activated. To
change the display timeout interval,
refer to Section 2.17.9.
Analyzer won’t respond when a
control panel key is pressed.
a. Microprocessor needs reset.
Remove analyzer’s top panel and
press the Reset button shown in
Figure 4-6.
b. Membrane switch defective.
Return to Bacharach for
replacement.