GA-40 T plus OPERATION
23
G Hu - Heating value of the fuel
G UNIT - unit in which the fuel is measured
œ DEFINE FUEL
1 NAME
2 CO2max [%] 15.8
3 A1 0.500
4 B 0.007
5 ALPHA 33
6 O2R [%] 3
7 VAtr [m3/l] 5
8 Hu [m3/l] 7
9 UNIT kg
5.2.5.2 Average Time
To select the average time proceed as follows:
 and à - move the cursor to Average Time
ENTER - select Average Time editing
0 to 9 - select the time ( 1 to 60 )
ENTER - confirm the number
 and à - select min or sec
ENTER - confirm the selection
5.2.5.3 Reference Oxygen parameter
To select the reference oxygen proceed as follows:
 and à - move the cursor to O2R
ENTER - switches between a pre-programmed standard value <NORM> and a freely
programmable one, called the manual value.
0 to 9 - if "manual" was selected, a number from 1 to 20 can be entered
ENTER - confirms and ends editing
5.2.5.4 Boiler Parameter
This option enables the user to have the relevant boiler parameters printed on the Report printout for exact
documentation.
 and à - move the cursor to BOILER-PARAMETER
ENTER - select with YES/NO whether or not the boiler parameters shall appear on the
printout, or be stored in a Report
Á - if YES was selected, this key moves to the Boiler Parameter menu. The boiler
parameters can be entered in the screen shown below: