OPERATION GA-40 T plus
30
œ FLOW MEASUREMENT
0 m/s
C - ZEROING
1 CORR. FACTOR : 0.10
C - Sets the zero point for the measurement
5.8 Analogue outputs (option)
The screen ANALOGUE OUTPUTS is reached from the main menu as follows:
 and à - move the cursor to ANALOGUE OUTPUTS option
Á - select the option, the following screen appears:
œ ANALOGUE OUTPUTS
1 CHANN. 1 : Tgas
2 FROM.....: 0°C
3 TO.......: 1000°C
5 CHANN. 2 : Tamb
6 FROM.....: 0°C
7 TO.......: 50°C
9 RANGE....: 0..20mA
For each channel it is possible to choose the variable and the range of the display separately.
The line RANGE enables the output to be switched between 0...20 mA and 4...20 mA.
5.9 Options
The OPTIONS can be reached from the MENU as follows:
 and à - move the cursor to OPTIONS
Á - select option. The following screen appears
œ
OPTIONS
1 DEVICE INFO
2 SENSORS
3 SETTINGS
4 CONTROL LIST
6 "madur" SERVICE