GA-40 T plus OPERATION
33
ENTER - default setting: Pt 500. The ambient temperature can either be measured with an
internal sensor placed in the probe plug (Pt500), or with an external sensor (Pt500),
connected to the TEMP1 (T3 EXT.) socket.
5.9.2.3.3 External thermocouple- - T1 EXT and T2 EXT
ENTER - Select the desired temperature sensor. The T1 EXT and T2 EXT inputs can be
defined as Ni-CrNi, Pt-RhPt or Fe-CuNi thermocouples
5.9.2.3.4 External resistance thermometer - T3 EXT and T4 EXT
ENTER - Select the desired temperature sensor. The inputs T3 EXT and T4 EXT can be
defined Pt100, Pt500 or KTY-2k thermistors.
5.9.2.3.5 External Voltage/Current inputs - U/I1 EXT and U/I2 EXT
Select the desired signal in the respective line by pressing [ENTER]
ENTER - Select the desired input signal . On both of the sockets U/I1 and U/I2 either current
(0-10 mA) or voltage (-10V - +10V) can be measured.
5.9.2.4 Sensor calibration
Besides O
2
every electrochemical sensor needs to be calibrated with calibration gas. To calibrate the sensors
they are fed one by one with calibration gas of a known concentration. After a few minutes the sensor
supplies a stabilised signal which can be stored as calibration value. The order of calibration does not effect
the calibration.
{{ Any calibration screen value change is followed by loss of the original
calibration settings. Thus CALIBRATION option should be selected only when
it is to be carried out.
The standard gas concentration should be similar to that usually met during measurement and it should not be
less than 10% of the required measuring range of a given sensor
The time the sensor is fed with standard gas should not be less than 4 min. After each calibration the sensors
should be flooded with ambient air for at least 15 minutes.
Calibration procedure
 and à - move the cursor to CALIBRATION W. GAS
Á - select the option
œ
ATTENTION
CALIBRATION!
ARE YOU
SURE
ENTER - CONTINUE
ENTER - carry on with the calibration
À - abort the calibration, move back to SENSORS