PCA
3 Manual
28 0024-9472 Rev 3
5. Repeat Step 4 until the values for month, day, and year have been set.
6. Press ENT to save the selection and re-display the SETUP MENU, or
press ESC to abort this procedure and retain the old date values.
NOTE:
The real time clock is powered by the main
batteries and is maintained by a coin cell battery on the
Main PCB in the absence of batteries. Bacharach
recommends changing the coin cell (P/N 0204-0020) every
5 years.
Figure 4-6. Date Setup
4.8. Time Setup
The time is stored in the format: hh:mm:ss AM/PM. Its value is part of the date
and time stamp that is saved along with each combustion test record. Set the
analyzer’s internal clock to the current time as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight TIME, and then press ENT to display
the TIME MENU.
4. First use the
◄►
buttons to move the cursor across the screen until it is
over the digit to be changed, and then press the ▲▼ buttons until the
desired value is displayed.
5. Repeat Step 4 until values for hour, minute, and meridian are set.
NOTE: The value for seconds cannot be entered, but is
displayed and stored as part of the combustion test record.
6. Press ENT to save the displayed time values and re-display the SETUP
MENU, or press ESC to abort and retain the old time values.