
PCA
3 Manual
0024-9472 Rev 3 73
and then press ENT to calibrate the SO
2
Measured value to that of the
applied value. The message “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen being re-displayed.
NOTE: If the sensor’s output is low, but still usable, then
the message “Good Calibration WARNING Low Sensor”
will appear. The sensor will now be marked as being Low
in the DIAGNOSTICS screen.
NOTE: If the sensor’s output is too low to be usable, then
the message “Bad Calibration Sensor End of Life, Entry
Not Saved” will appear followed
LIST screen being re-
displayed. The sensor will now be
marked as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove the gas cylinder.
6.9. NO Sensor Calibration
This procedure spans the optional nitric oxide sensor to a known gas
level.
6.9.1. Material Required
•
Calibration Kit (P/N 0024-7059)
•
Gas Cylinder ( 250 ppm NO in Nitrogen, P/N 0024-1156)
6.9.2. Procedure
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 62).
2. Use the ▲▼ buttons to highlight NO, and then press ENT to display the
CALIBRATE NO screen.
NOTE:
“Measured” is the current NO reading, while
“Applied” is a known NO level that will be applied for
calibration purposes.
3. Attach a 250 ppm NO cylinder to the regulator of the calibration fixture
shown in Figure 6-12 on page 76.
4. Use the ▲▼ and
◄►
buttons to enter an applied value that exactly
equals the concentration stamped on the NO cylinder.
NOTE: The calibration range is from 9 to 900 ppm. An
attempt to calibrate outside this range will cause the
message “Bad Calibration Wrong CAL Entry” to appear in
the following step.
Commentaires sur ces manuels