PCA
3 Manual
90 0024-9472 Rev 3
O
2
Sensor
Missing
Oxygen sensor not installed. Install oxygen sensor. Refer to
Sensor Replacement on page 84.
BAD Sensors
The oxygen sensor’s output is too low and can not be calibrated
in the instrument, signifying that the sensor is depleted and
needs to be replaced. Refer to Sensor Replacement on page 84.
No B-Smart
Sensors Installed
There are no B-Smart sensor(s) installed. Refer to Sensor
Replacement on page 84.
Low Sensors
O
2
, CO
LOW
, CO
HIGH
, NO, NO
2
, or SO
2
sensor output(s) were low,
but still usable. Sensor(s) may need to be replaced in the near
future. Message will indicate which sensor(s) are in warning.
T-STK
Disconnected
The probe’s thermocouple is not connected to the analyzer’s
T-STACK connector. Plug the probe thermocouple plug into the
T-STACK connector at the bottom of the instrument.
Warm-up Sensor
Error
CO
LOW
, CO
HIGH
, NO, NO
2
, or SO
2
Gas sensor(s) were not zeroed
at warm-up because of high output. Run instrument on fresh air
then restart instrument to re-zero sensor(s). If message persists,
sensor(s) may need to be replaced. Message will indicate which
sensor(s) are in error.
•
Stack or Air temperature channel is measuring temperature
outside the range of -4 to 212 °F at startup. Make sure that
the Stack and Air thermocouples are sampling ambient
room air within the temperature range at startup. Message
will indicate which channel(s) are in error.
•
Pressure sensor is measuring pressure outside the range of
+/- 3 inches of water column at startup. Ensure that the
analyzer is sampling atmospheric pressure and restart.
•
The analyzer was turned on with the probe sampling flue
gas. Move the probe to fresh air and restart the analyzer.
8.6. Replacement Parts
Item Description Part No.
2 LCD Module 0024-1520
3 Rear Case 0024-1381
4 Top Case 0024-1524
5 Battery Cover 0024-1383
6 Window 0024-1526
7 LCD Frame 0024-1525
8 Keypad 0024-1387
9 Chassis 0024-1388
10 Lens, Infrared 0024-1391
11 Replacement Pump, Sample Gas 0024-1547
12 Replacement Pump, Purge 0024-1548
13 Sensor Retainer 0024-1418
14 Replacement Hose Connector Plate (complete 0024-1480