PCA
3 Manual
68 0024-9472 Rev 3
1. Set thermocouple simulator to room temperature and plug its output into
the T-AIR connector located at the bottom of the analyzer.
Alternatively:
Plug the probe’s thermocouple into the T-AIR connector
located at the bottom of the analyzer. DO NOT attach the probe’s gas
hose to the analyzer’s GAS port; otherwise water will be drawn into the
analyzer!
2. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 62).
3. Use the ▲▼ buttons to highlight T-Air, and then press ENT to display
the CALIBRATE TA-ZERO screen.
NOTE: “Measured” is the current t
emperature reading,
while “Applied” is a known temperature that will be applied
for calibration purposes.
4. Set thermocouple simulator to 32 °F (0 °C), and then use the ▲▼ and
◄► buttons to enter an applied value that exactly equals the setting of
the simulator.
Alternatively:
Submerge probe tip into an ice-water bath with a
thermometer, wait several minutes, and then use the ▲▼ and ◄►
buttons to enter an applied value that exactly equals the thermometer
reading.
NOTE: The calibration range is from 32 to 41 °F (0 to
5 °C). An attempt to calibrate outside this range will cause
the message “Applied Value High” (or Low) to appear at
the bottom of the screen.
5. Wait until the measured reading stabilizes, and then press ENT to
calibrate the TA-Zero Measured value to that of the applied value; after
which the message “Good Calibration” should briefly appear followed by
the CALIBRATE TA-SPAN screen.
Figure 6-5. T-Air Zero Procedure