PCA
3 Manual
36 0024-9472 Rev 3
4.16. Language
Information on the display screen can be shown in English, French, or Spanish.
Select the desired language as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight LANGUAGE SELECTION, and then
press ENT to display The Language Selection Menu.
4. Use the ▲▼ buttons to highlight the desired language, then press ENT
to activate the selection and re-display the Setup Menu.
Figure 4-18. Selecting a Language
4.17. Cal Reminder Period
The analyzer can be set to indicate a calibration reminder during the 60 second
warm-up period. Calibration reminders can be preset to occur never or at 6, 8,
10, 12, or 15 months after the last calibration. When the preset period is
exceeded the instrument will display the reminder, and how long since the
sensors were last calibrated. The reminder will be displayed at the end of the 60
second warm-up period. If a calibration reminder is displayed the operator can
press the RUN/HOLD key to move to the Run/Hold screen for normal operation.
Regular calibration periods of 6 months to a year for all gas sensors (except
Oxygen) are recommended. Set the calibration reminder period as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight CAL REMINDER PERIOD, and then
press ENT to display CAL Reminder Period Menu.
4. Use the ▲▼ buttons to highlight the desired CAL reminder period, and
then press the ENT key to activate the selection and re-display the
Setup Menu.