PCA
3 Manual
0024-9472 Rev 3 51
Logging Intervals and Durations
Logging Intervals Logging Durations
1 second 5 minutes
5 second 10 minutes
10 second 15 minutes
15 second 30 minutes
30 second 1 hour
1 minute 2 hours
2 minutes 5 hours
5 minutes 10 hours
10 minutes 24 hours
48 hours
5.12.2. Turning On Data Logging
IMPORTANT: Before turning ON data logging and starting the data
logging process, the analyzer should already be set up to perform a
combustion test (see Performing a Combustion Test section on
page 43).
Turn ON data logging as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight LOGGING, and then press ENT to
display the LOGGING screen.
NOTE: The amount of memory available for storing new
data is displayed in the LOGGING screen (maximum of
500 locations). If previous logging sessions are stored, and
additional memory is needed, clear the logging memory
(see Clearing Memory section on page 56).
NOTE:
At no time will new logging data over write old
data. Logging will stop when available memory is filled.
4. Use the ▲▼ buttons to highlight YES (turn logging ON).
5. Press ENT to make the selection and display the LOGGING INTERVAL
screen.