PCA
3 Manual
0024-9472 Rev 3 77
NOTE: The calibration range is from 250 to 11,000 ppm.
An attempt to calibrate outside this range will cause the
message “Bad Calibration Wrong CAL Entry” to appear in
the following step.
Figure 6-13. CO
HIGH
Sensor Calibration
5. Adjust regulator for a flow meter indication of approximately 2 SCFH.
Wait until the measured reading stabilizes (approximately 3 minutes),
and then press ENT to calibrate the CO Measured value to that of the
applied value. The message “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen being re-displayed.
NOTE: If the sensor’s output is low, but still usable, then
the message “Good Calibration WARNING Low Sensor”
will appear. The sensor will now be marked as being Low
in the DIAGNOSTICS screen.
If the sensor’s output is too low to be usable, then the
message “Bad Calibration Sensor End of Life, Entry Not
Saved” will appear followed by the CALIBRATION LIST
screen being re-displayed. The sensor will now be marked
as being BAD in the DIAGNOSTICS screen.
6. Turn OFF regulator and remove the gas cylinder.